Frequently Asked Questions
Q. How
do I set the oven flow on a 491 MB base instrument?
A. Place
the instrument in the
Span position with the
Flow
toggle button pushed to display the flow amount on the panel meter.
Dial the
Dilution
Flow Adjust knob counterclockwise until finger tight. The
remaining flow amount
seen on the display will be the oven flow. Turn the Oven Flow Adjust
knob clockwise to increase the flow until it reads at least
0.10 slpm. This sets the required 100 cc/min of oven flow. Now
turn the
Dilution Flow Adjust knob clockwise to
increase flow to your calculated level. When
you switch to the
Zero position you will be able to
read the
Dilution Flow
setting only because the
Oven Flow is vented and
not metered.
In the
Standby position,
Oven
Flow is also vented
and dilution flow is stopped. In
Standby there will
be no flow
displayed and the panel will read “0.00”.
[
back to list of FAQs]