Frequently Asked Questions
Q. What
is the Ko factor that I use in concentration calculations?
A. The
conversion from nl/min to ng/min is easy when using the Ko factor.
Ko
= 22.4 / MW
Where:
Ko =
desired conversion factor
MW
= molecular weight of the pure analyte.
Then, E1 = E2 / Ko
Where:
E1 = desired
emission rate in ng/min
E2 = emission
rate in nl/min
Ko = conversion
factor as specified above
[