Q
. How do I calculate the concentration I need when using a disposable tube?

A. For a KIN-TEK disposable tube, use the following formula for concentration in ppmv:

     C = (E * Ko) / F

     Where:

          C = desired concentration in ppmv.
          E = certified emission rate (ng/min) as reported on
                 your tube certificate.
          Ko = value that corresponds to the compound-specific
                    Ko on your tube certificate.
          F = rate of dilution flow you are using (cc/min)