FAQ #1
Q.
How do I calculate the
concentration I need when using a disposable tube?
A. For a
KIN-TEK
disposable tube, use the following formula for
concentration in ppmv:
C = (E * Ko) / F
Where:
C = desired
concentration in ppmv.
E
= certified emission rate (ng/min) as reported on
your tube certificate.
Ko
= value
that corresponds to the compound-specific
Ko
on your tube certificate.
F
= rate of
dilution flow you are using (cc/min)