Model 491M Gas Standards Generator

FlexStream automated gas standards generator

The 491M gas standards generator provides:
Industry-standard for over 20 years
Permeation tube oven with wide temperature range controlled to 0.1 °C
Primary dilution mass flow rate
Standby, zero and span modes
Single-oven module can create mixtures with up to 8 components
Expandable

 

 

  • Overview
  • Specifications
  • Compare with FlexStream

The 491M is ideally suited for creating trace concentration - ppm, ppb, and pptr - mixtures. The 491M uses KIN-TEK Trace Source™ disposable, refillable and diffusion permeation tubes for over 500 chemicals.

The 491M is composed of one or more of the following manually-operated modules, depending on application requirements:

  • 491MB Base Module (shown above): The "Base" module is required in every 491M instrument. This module provides manual front panel (a) oven temperature setpoint and primary dilution flow rate adjustment controls, (b) standby, zero and span selection, and (c) permeation tube oven temperature and primary dilution mass flow rate indication. [brochure]

  • 491M-PM Auxiliary Permeation Oven Module (PM): Additional permeation tube ovens allow generation of a multi-component gas stream for cases in which (a) the tubes must be maintained at different temperatures, (b) differing permeation technologies (disposable tubes vs. refillable tubes) are required, (c) more components are needed than can be held in a single oven, or (d) for applications in which various components will be switched in and out of the gas stream. Up to 5 PM oven modules can be added wherein each oven can hold up to eight disposable permeation tubes or one factory-refillable tube. Any combination of ovens can be kept on standby to keep various tubes at temperature while not adding their emission to the output gas stream. [brochure]

  • 491M-SD Secondary Dilution Module (SD): The SD Module extends the range of concentration attained from permeation tubes by providing an additional stage of dilution of the gas stream. Very low concentrations - ppb or ppt - can be generated depending on the emission rate of the tubes and the flow rate range of the primary and secondary dilution flow controllers. Overall dilution ratios of 10,000:1 can be achieved. [brochure]

  • 491M-GF Gas Feed Module (GF): Most applications can use liquid-filled permeation tubes. However, some compounds (e.g., NO, CO, O2, etc.) have very high vapor pressures and require a special mechanism to achieve desired concentration levels. The Gas Feed Module adapts an adjacent 491MB Base oven module or auxiliary 491-PM oven module and allows the customer to refill a "customer-refillable" gas-fed permeation tube. These are special tubes, typically made of stainless steel. Prior to shipment, KIN-TEK determines the tube membrane's certified emission rate for one or more components. In use, the customer typically supplies component gas from a cylinder to initially fill the tube and to periodically recharge it. [brochure]

  • 491M-DGB Direct Gas Blending Module (DGB): The DGB Module allows input of higher concentration component gas from a cylinder into the gas stream.The 491M gas standards generator then dilutes the mixture to produce a desired concentration. Dilutions from 1.5:1 to 200:1 are possible. [brochure]

  • 491M-HG Humidified Gas Module (HG): The Humidification Module allows addition of humidity to the gas stream. Up to 95% relative humidity can be added, depending on the overall dilution flow rate required to achieve the target concentration at the target relative humidity. [brochure]

  • 491M-IM Interface Module (IM): The Interface Module contains gas stream interfacing components needed to assure constant gas stream output conditions independent of variations in flow and pressure. The Interface Module is ideal for applications where flow capacity of end use is limited (e.g., filling of gas bags), or when back pressure is required to introduce the output gas stream sample to the device under test. Controls are provided on the front panel for adjusting the output pressure. [brochure]

  • 491M Modular Gas Standards Generator (shown below): The 491MB Base module and any additional modules can be packaged in stand-alone carrying cases or packaged in one or two 3-module expansion cases as shown below (unused module slots in the expansion case are filled with blank panels). [brochure]

491M Modular Gas Standards Generator

491M Modular System

The example 491M instrument above contains a 491MB Base Module, 491M-PM Auxiliary Permeation Oven Module and 491M-IM Interface Module.

491M features:

  • Industry standard for over 20 years
  • Complete, integrated, ready-to-use turnkey system
  • Manually operated via front panel controls
  • Direct readout of oven temperature and mass flow rate
  • Modes of Operation: standby, zero and span
  • Designed for expandability
  • Flow path designed for maximum flexibility with minimum error
  • Flow path suitable for reactive gases - mixture contacts only glass, Teflon® and stainless steel (other materials available)
  • Optional component flow path materials available (sulfinert-coated, all stainless, etc.)
  • Accepts disposable permeation tubes, diffusion tubes, ultra-high rate liquid filled tubes, wafer tubes, and prefilled gas fed permeation tubes
  • Accepts up to 8 disposable tubes per oven with maximum 6 inch length x 1/4 inch diameter (KIN-TEK HRT, SRT and EL tubes)
  • Accepts one of KIN-TEK refillable LFH, ULED and 57 Series tubes per oven

Please scroll up and click the "Specifications" tab above to see more features and specifications of the 491M.

Principle of Operation 491MB
Base Module
     Dilution of constant component vapor emission from permeation or
     diffusion tubes
     Delivery of precise, adjustable, low concentration dynamically blended
     gas mixtures. Trace concentrations - ppmv, ppbv, pptv
     Manual front panel controls for oven temperature and dilution flow rate
     adjustment and standby, zero and span selection
   
Permeation Tube Oven  
     Construction - high thermal mass oven with electronic PID control Vertical orientation
     Oven temperature range (heat only) 30-150 °C
     NIST-traceable oven temperature
     Oven setpoint resolution and control ±0.1 °C
     Temperature digital indication and adjust via front panel
   
Diluent Gas (customer-supplied)  
     Uses inert, non-reactive, non-hazardous dry gas (Nitrogen, zero air,
     Argon, etc.)
     Mass flow meter subsystem is factory-calibrated in Nitrogen; contact factory
     for calibration in other type gas
       
Primary Dilution Mass Flow Controller
     Dilution flow rate set by manual valve on front panel; measured by mass flow meter
     Standard dilution flow rate range 0.25 to 5 lpm
     Optional flow rate ranges available (0.5-10, 0.1-1, 0.1-0.5 lpm)
     Mass flow rate accuracy ±1% FS
     Mass flow rate indication via front panel
     Mass flow calibration method Analog zero/span adjust
   
Accepts Vapor Emissions from the Following Devices*  
     Liquid-filled disposable tubes up to 15 cm in length (maximum of 8 tubes)
     Liquid-filled high rate (LFH) permeation tubes (maximum of 1 tube)
     Factory pre-filled gas-fed permeation tubes (maximum of 1 tube)
     Diffusion tubes (maximum of 4 tubes)
   
*Note: A single oven can hold only one of the above tube types at a time  
   
Standard Component Flow Path  
     Component contacts only glass, Teflon® and stainless steel
     Glass permeation tube holder
     Suitable for reactive component gases
   
Optional Component Flow Path (specified at time of order)  
     Heated output lines
     Stainless steel
     Electro-polished stainless steel
     Sulfinert-coated stainless steel
     Teflon only
     Teflon permeation tube holder
   
Standard Modes of Operation  
     Standby (oven venting at approx 75 cc/min flow rate)
     Zero (with user-specified flow rate)
     Span-by-flow; flow rate set by manual valve on front panel and measured
     by mass flow meter with front panel digital indication
   
Output Concentration Range  
     Below 1 ppb to over 1000 ppm depending on tube emission rate
     and dilution flow rate; low ppb and ppt concentrations with secondary dilution
   
Expandability  
     Additional permeation ovens (via up to five 491M-PM modules)
     Customer-refillable gas-fed permeation oven option (via 491M-GF modules)
     Secondary diluition flow rate option (via 491M-SD module); other ranges available

0.5 to 5 lpm

     Overall output gas stream dilution ratio with both primary and secondary dilution Up to 10,000 to 1
     Direct gas blending module option (via 491M-DGB)
     Humidified gas module option (via 491M-HG)
     Gas stream output Interface Module option (via 491M-IM module);
     If 491M-SD secondary dilution module is installed, subtract 10 psig (69 kPa)
ambient to 40 psig (276 kPa)
   
Instrument Housing (specified when ordering)  
     Carrying case (standard)
     Expansion chassis (option); desktop or rack-mount; 2 cases maximum 3 modules per case
     Maximum number of modules in a single 491M system 6
   
Power Requirements (specified when ordering)  
     115 VAC, 50/60 Hz
     230 VAC, 50/60 Hz
   
Dimensions and Weight  
     Dimensions (Base module as packaged in carrying case); approximate
7.5 W x 13.5 H x 20 D inch
18.4 W x 34.3 H x 50.8 D cm
     Weight (Base module as packaged in carrying case); approximate 32 lb (14.5 kg)
   
Certifications  
     European CE Mark (230 VAC, 50 Hz): 491MB (Base), 491M-PM, 491M-SD,
     491M-HG, 491M-GF, 491M-DGB
     UL (site-specific)
   
Speciality Applications (Please contact the Factory)  
     Special diluent (carrier) gases other than inert gases
     Trace moisture
     Chemical compatibility issues
     Heat tracing requirements
     Other

 

Principle of Operation
491MB
Base Module
FlexStream Base Module
     Dilution of constant component vapor emission from
     permeation or diffusion tubes
     Delivery of precise, adjustable, low concentration
     dynamically blended gas mixtures. Trace concentrations -
     ppmv, ppbv, pptv
     Oven temperature and dilution flow rate adjustment and
     standby, zero, and span selection
Manual controls Touchscreen
     External PC monitoring and control via industry-standard
     Modbus® RTU communications protocol and Ethernet or serial
     RS-232 connectivity
     
Permeation Tube Oven    
     Construction - high thermal mass oven with electronic PID control Vertical orientation Vertical orientation
     Oven temperature range (heat only) 30-150 °C 30-150 °C
     NIST-traceable oven temperature
     Oven setpoint resolution and control ±0.1 °C ±0.01 °C
     Temperature digital indication and adjust via front panel
     Over-temperature detection with audible alarm and oven
     power removal
     Temperature digital indication and adjust via computer
      interface
     
Diluent Gas (customer-supplied)    
     Carrier flow rate across oven manually-set fixed
     Uses inert, non-reactive, non-hazardous dry gas
     (Nitrogen, zero air, Argon, etc.)
     Mass flow subsystem is factory-calibrated in Nitrogen;
     contact factory for calibration in other type of gas
     Accepts a gas correction factor (GCF) for most gases to
     auto-adjust the dilution mass flow rate when a diluent gas
     other than Nitrogen is used
     Using the supplied FlexLink™ PC-based software (FlexStream
     only), the mass flow controller subsystem can be easily calibrated
     with up to 10 points in other diluent gases (Customer supplies the
     flow standard)
     
Primary Dilution Flow Rate    
     Dilution flow rate type Mass flow meter Mass flow meter and controller
     Standard dilution flow rate range 0.25 to 5 lpm 0.25 to 5 lpm
     Optional flow rate ranges available 10, 1, 0.5 lpm 10, 1, 0.5 lpm
     Flow rate accuracy ±1% FS Lesser of ±1.5% Reading and ±1% FS
     Flow rate calibration method Analog zero/span adjust Piece-wise linear fit of up to 10 calibration points
     Flow rate indication via front panel
     Flow rate control via front panel Flow rate set by manual valve from front panel Flow rate set from touch screen target flow or target concentration
     Flow rate digital indication and control via computer
     interface
     Flow rate can be auto-set based on target concentration
     
Accepts Vapor Emissions from the Following Devices*    
     Liquid-filled disposable tubes up to 15 cm in length
     (maximum of 8 tubes)
     Liquid-filled high rate (LFH) permeation tubes (maximum
     of 1 tube)
     Factory pre-filled gas-fed permeation tubes (maximum
     of 1 tube)
     Diffusion tubes (maximum of 4 tubes)
     
*Note: A single oven can hold only one of the above tube types at a time    
     
Standard Component Flow Path    
     Component contacts only glass, Teflon® and stainless
     steel
     Glass permeation tube holder
     Suitable for reactive component gases
     
Optional Component Flow Path    
     Heated output lines
     Stainless steel
     Electro-polished stainless steel
     Sulfinert-coated stainless steel
     Teflon only
     Teflon permeation tube holder
     
Standard Modes of Operation    
     Standby
     Zero
     Span-by-flow Manual valve set and digital indication Touchscreen flow setpoint and digital indication
     Span-by-concentration (user-specified target concentration)
     Auto-calculates flow setpoint based on target
     concentration, tube emission rate and oven temperature

Manual
Calculation

     Automatically updates touchscreen with up to eight
     generated concentration values
     
Output Concentration Range    
     Below 1 ppb to over 1000 ppm depending on tube
     emission rate and dilution flow rate; low ppb and ppt
     with secondary dilution
     User selects units of ppm, ppb or ppt
     
Interfaces    
     Local human interface - color touch screen with virtual
     keypad
     Remote computer interfaces (standard Ethernet & RS-232)
     Communications Protocol Modbus RTU
     
PC-based Software    
     KIN-TEK FlexLink™ software CD shipped with instrument
     FlexStream Labview® Modbus driver on FlexLink™ CD
     LabView®-based starter software with source available
     
Expandability    
     Additional permeation ovens
     Customer-refillable gas-fed permeation oven option
     Secondary diluition flow rate option; other ranges
     available
0.5 to 5 lpm

0.5 to 5 lpm

     Overall output gas stream dilution ratio with both primary
     and secondary dilution
Up to 10,000:1 Up to 10,000:1
     Humidified gas module option Contact factory
     Gas stream output Interface module option;
     If secondary dilution module is installed, subtract 10 psig (69 kPa)
ambient to 40 psig (276 kPa) ambient to 30 psig (207 kPa)
     
Instrument Housing (specified when ordering)    
     Carrying case (standard)
     Expansion chassis (option); desktop or rack-mount;
     2 cases maximum
3 modules per case 3 modules per case
     Maximum number of modules in a single system 6 6
     
Power Requirements (specified when ordering)    
     115 VAC, 50/60 Hz
     230 VAC, 50/60 Hz
     
Dimensions and Weight    
     Dimensions (Base module in carrying case) Similar Similar
     Weight (Base module in carrying case) Similar Similar
     
Certifications    
     European CE Mark (230 VAC, 50 Hz)
     NRTL (site-specific) not tested
     CSA (site-specific) not tested
     UL (site-specific) not tested
     
Speciality Applications (Please contact the Factory)    
     Special diluent (carrier) gases other than inert gases
     Trace moisture
     Chemical compatibility issues
     Heat tracing requirements
     Other

 